AC-DC transfer comparator for the calibration of thermal voltage converters against josephson alternating voltage standards
Abstract
This paper describes a new measurement system for the comparison of alternating voltage standards based on the Josephson effect with thermal voltage converters. The target is to achieve an uncertainty of 0.1 muV/V at frequencies up to 10 kHz. The present work achieves an uncertainty of 0.2 muV/V up to 2 kHz.
Year
DOI
Venue
2003
10.1109/TIM.2003.811671
IEEE T. Instrumentation and Measurement
Keywords
Field
DocType
Josephson effect,calibration,comparators (circuits),convertors,measurement standards,measurement uncertainty,voltage measurement,2 kHz,AC-DC transfer comparator,Josephson alternating voltage standard,calibration,measurement uncertainty,thermal voltage converter
Comparator,Josephson effect,System of measurement,Voltage,Voltage reference,Measurement uncertainty,Electronic engineering,Converters,Electrical engineering,Calibration,Mathematics
Journal
Volume
Issue
ISSN
52
2
0018-9456
Citations 
PageRank 
References 
2
0.53
3
Authors
3
Name
Order
Citations
PageRank
I. Budovsky1175.07
H. Sasaki220.87
P. Coogan320.53