Title
Progress Toward a 1 V Pulse-Driven AC Josephson Voltage Standard
Abstract
We present a new record root mean square (rms) output voltage of 275 mV, which is a 25% improvement over the maximum that is achieved with previous ac Josephson voltage standard (ACJVS) circuits. We demonstrate the operating margins for these circuits and use them to measure the harmonic distortion of a commercial digitizer. Having exceeded the threshold of 125 mV rms for a single array of Josephson junctions, we propose and discuss the features of an eight-array circuit that is capable of achieving 1 V rms. We investigate the use of a resistive divider to extend the ACJVS voltage accuracy to higher voltages. By the use of a switched-input measurement technique, an integrating sampling digital voltmeter, a resistive voltage divider, and ACJVS synthesized sine waves as reference voltages, we characterize the stability of a commercial calibration source for a few voltages up to 2.7 V.
Year
DOI
Venue
2009
10.1109/TIM.2008.2007019
IEEE T. Instrumentation and Measurement
Keywords
DocType
Volume
josephson junctions,sampling methods,josephson junction,voltage,quantization,superconductors,voltmeters,stability,root mean square,calibration,harmonic distortion
Journal
58
Issue
ISSN
Citations 
4
0018-9456
5
PageRank 
References 
Authors
0.84
4
6
Name
Order
Citations
PageRank
Samuel P. Benz110620.78
Paul D. Dresselhaus29817.82
Alain Rüfenacht39615.03
Norman F. Bergren450.84
Joseph R. Kinard5153.55
Regis Pinheiro Landim6185.18