Abstract | ||
---|---|---|
A method is presented to estimate the acquisition geometry of a pinhole single photon emission computed tomography (SPECT) camera with a circular detector orbit. This information is needed for the reconstruction of tomographic images. The calibration uses the point source projection locations of a tomographic acquisition of three point sources located at known distances from each other. It is show... |
Year | DOI | Venue |
---|---|---|
2003 | 10.1109/TMI.2003.812258 | IEEE Transactions on Medical Imaging |
Keywords | Field | DocType |
Image reconstruction,Calibration,Imaging phantoms,Detectors,Extraterrestrial measurements,Computational geometry,Single photon emission computed tomography,Cameras,Algorithm design and analysis,Least squares approximation | Least squares,Single-photon emission computed tomography,Iterative reconstruction,Computer vision,Imaging phantom,Data acquisition,Point source,Artificial intelligence,Geometry,Detector,Calibration,Physics | Journal |