Title
An Implantable Long-Term Bladder Urine Pressure Measurement System With A 1-Atm Canceling Instrumentation Amplifier
Abstract
An implantable system for long-term bladder urine pressure measurement system is presented. Not only is the design cost reduced, but also the reliability is enhanced by using a 1-atm canceling sensing IA (instrumentation amplifier). Because the urine pressure inside the bladder does not vary drastically, both the sleeping and working modes are required in order to save the battery power for long-term observation. The IA amplifies the signal sensed by the pressure sensor, which is then fed into the following ADC (analog-to-digital converter). Owing to the intrinsic 1-atm pressure (one atmospheric pressure) existing inside the bladder, the IA must be able to cancel such a pressure from the signal picked up by the pressure sensor to keep the required linearity and the resolution for pressure measurement of the bladder urine. The pressure range of the proposed system is found out to be 14.7 similar to 19.7 Psi, which covers the range of all of the known unusual bladder syndromes or complications.
Year
DOI
Venue
2007
10.1109/ISCAS.2007.377939
2007 IEEE INTERNATIONAL SYMPOSIUM ON CIRCUITS AND SYSTEMS, VOLS 1-11
Keywords
Field
DocType
urine pressure, implantable, bladder, IA, linearity
Computer science,Linearity,Pressure measurement,Instrumentation amplifier,Analog-to-digital converter,Atmospheric pressure,Electronic engineering,Pressure sensor,Battery (electricity),Atmosphere (unit)
Conference
ISSN
Citations 
PageRank 
0271-4302
2
2.63
References 
Authors
0
8
Name
Order
Citations
PageRank
Chua-chin Wang1474107.39
Chi-chun Huang210117.97
Jian-sing Liou3135.47
Yan-Jhin Ciou4165.63
I-Yu Huang5146.85
Chih-Peng Li633742.52
Yun-chin Lee722.63
Wen-jen Wu843.07