Title
Spatiotemporal Evaluation Of Human Colon Motility Using Three-Axis Fluxgates And Magnetic Markers
Abstract
An alternative method to study the mechanical activity of the human colon in fasting and postprandial states is presented. The method is based on measurements of the magnetic fields produced by a magnetic marker, a small cylindrical NdBFe magnet, when it was ingested by the subjects. A portable magnetic probe, consisting of two digital three-axis fluxgate magnetometers, arranged in a first-order electronic gradiometer, was implemented for this research. Measurements were taken in 16 healthy male subjects. Contractile activity frequency measurements were taken along the colon length, including the ascending, transverse and descending sections, as well as the rectal sigmoidal section. Values for the contractile activity frequency of 2-5 cycles min(-1) were measured. The set-up is simple, low-cost and suitable for use in an unshielded environment.
Year
DOI
Venue
2005
10.1007/BF02430947
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING
Keywords
Field
DocType
magnetic marker, contractile activity frequencies, colon motility
Magnetic field,Transverse plane,Fluxgate compass,Gradiometer,Magnetometer,Magnet,Electronic engineering,Motility,Mathematics
Journal
Volume
Issue
ISSN
43
6
0140-0118
Citations 
PageRank 
References 
1
0.41
0
Authors
6
Name
Order
Citations
PageRank
T Córdova-Fraga111.76
A A O Carneiro210.41
D B de Araujo331.28
R B Oliveira410.41
M Sosa510.75
O Baffa610.75