Title
Model-Based Development of the Generic PCA Infusion Pump User Interface Prototype in PVS
Abstract
A realistic user interface is rigorously developed for the US Food and Drug Administration (FDA) Generic Patient Controlled Analgesia (GPCA) pump prototype. The GPCA pump prototype is intended as a realistic workbench for trialling development methods and techniques for improving the safety of such devices. A model-based approach based on the use of formal methods is illustrated and implemented within the Prototype Verification System (PVS) verification system. The user interface behaviour is formally specified as an executable PVS model. The specification is verified with the PVS theorem prover against relevant safety requirements provided by the FDA for the GPCA pump. The same specification is automatically translated into executable code through the PVS code generator, and hence a high fidelity prototype is then developed that incorporates the generated executable code.
Year
DOI
Venue
2013
10.1007/978-3-642-40793-2_21
SAFECOMP
Keywords
Field
DocType
model based development,formal methods
Prototype Verification System,Workbench,Computer science,Automated theorem proving,Model-based design,Code generation,Formal methods,User interface,Executable,Embedded system
Conference
Citations 
PageRank 
References 
11
0.64
16
Authors
6
Name
Order
Citations
PageRank
Paolo Masci118829.52
Anaheed Ayoub21048.15
Paul Curzon341640.23
Insup Lee44996413.64
Oleg Sokolsky52193154.94
Harold W. Thimbleby664090.12