Title
Calibration of Rogowski Coils With an Integrator at High Currents.
Abstract
This paper describes the calibration of Rogowski coils with an integrator for current measurement in the current range of up to 30 kA, at mains frequency. A current loop of ten turns with a current of 3 kA was used for calibration. The magnetic field generated in this way differs from the field of one turn placed at the center and passed with a current of 30 kA. The influence of the unevenness of ...
Year
DOI
Venue
2011
10.1109/TIM.2011.2126090
IEEE Transactions on Instrumentation and Measurement
Keywords
Field
DocType
Uncertainty,Current measurement,Calibration,Coils,Voltage measurement,Conductors,Inductance
Magnetic field,Inductance,Electrical conductor,Integrator,Measurement uncertainty,Electronic engineering,Current loop,Rogowski coil,Electrical engineering,Mathematics,Calibration
Journal
Volume
Issue
ISSN
60
7
0018-9456
Citations 
PageRank 
References 
7
1.20
3
Authors
4
Name
Order
Citations
PageRank
Karel Draxler1113.88
Renata Styblikova2113.88
Jan Hlavacek371.20
Radek Prochazka471.20