Title
A Broadside-Split-Ring Resonator-Based Coil for MRI at 7 T.
Abstract
A coil design termed as broadside-coupled loop (BCL) coil and based on the broadside-coupled split ring resonator (BC-SRR) is proposed as an alternative to a conventional loop design at 7T. The BCL coil has an inherent uniform current which assures the rotational symmetry of the radio-frequency field around the coil axis. A comparative analysis of the signal-to-noise ratio provided by BCL coils and conventional coils has been carried out by means of numerical simulations and experiments in a 7T whole body system.
Year
DOI
Venue
2013
10.1109/TMI.2013.2253488
IEEE Trans. Med. Imaging
Keywords
Field
DocType
numerical simulation,radio frequency,magnetic resonance imaging,capacitance,capacitors,meta materials,signal to noise ratio,comparative analysis,rotational symmetry,numerical analysis,computer simulation,metamaterials,mri
Broadside,Coil noise,Artificial intelligence,Search coil,Rotational symmetry,Computer vision,Electromagnetic coil,Acoustics,Numerical analysis,Split-ring resonator,Microstrip resonators,Nuclear magnetic resonance,Mathematics
Journal
Volume
Issue
ISSN
32
6
0278-0062
Citations 
PageRank 
References 
0
0.34
1
Authors
7
Name
Order
Citations
PageRank
Manuel J. Freire131.69
Marcos A. Lopez200.34
Florian Meise300.34
Jose M. Algarin410.73
Peter M Jakob5245.25
M. Bock600.68
Ricardo Marqués721.30