Title
Research on tipping calibration and exponent calibration based on LN2 for ground-based multi-channel microwave radiometer.
Abstract
Respect to ground-based microwave radiometer, the paper presents exponent calibration based on blackbody cooled by liquid nitrogen. In order to correct the calibration coefficients, this paper also presents tipping method at low optical depth (K band), which is different from common tipping method. It observes the clear-sky and clear-sky with noise injection to correct nonlinear factor and stability of noise. The calibration results show that using tipping method can achieve high resolution. © 2012 IEEE.
Year
DOI
Venue
2012
10.1109/IGARSS.2012.6350428
IGARSS
Keywords
Field
DocType
stability,liquid nitrogen,calibration,radiometers,temperature measurement,microwave radiometer,noise,ln2
Optical depth,K band,Nonlinear system,Exponent,Computer science,Remote sensing,Optics,Black-body radiation,Calibration,Microwave radiometer,Radiometer
Conference
Volume
Issue
ISSN
null
null
null
ISBN
Citations 
PageRank 
978-1-4673-1158-8
0
0.34
References 
Authors
1
3
Name
Order
Citations
PageRank
Jieying He1109.68
Sheng-Wei Zhang210616.04
Yu Zhang300.34