Title
Uncertainties of Multiport VNA S-Parameter Measurements Applying the GSOLT Calibration Method
Abstract
For the general short-open-load-thru (SOLT) calibration of the n-port vector network analyzer with n + 1 measurement channels, the sensitivity coefficients for the S-parameters of the n-port device under test are developed as functions of the deviations of the scattering parameters for the SOLT calibration standards. Using the concept of general node equation, a generalized formula for the S-parameter deviations with respect to the error terms has been deduced. In addition, expressions representing the deviations of error terms with respect to the non-ideal calibration standards are given by a series of matrix operations. Finally, after calculation of sensitivity coefficients, they can be used for establishing the type-B uncertainty budget for S-parameter measurements.
Year
DOI
Venue
2012
10.1109/TIM.2012.2210332
IEEE T. Instrumentation and Measurement
Keywords
Field
DocType
network analysers,measurement errors,calibration,measurement channel,s-parameters,general node equation,n-port vector network analyzer,multiport vector network analyzer (vna),measurement uncertainty,scattering parameter,general short-open-load-thru (gsolt) calibration,n-port device under test,gsolt calibration method,multiport vna s-parameter measurement,type-b uncertainty budget,general short-open-load-thru calibration method,sensitivity coefficient,s parameters
Network analyzer (electrical),Device under test,Expression (mathematics),Measurement uncertainty,Electronic engineering,Scattering parameters,Matrix multiplication,Mathematics,Calibration,Observational error
Journal
Volume
Issue
ISSN
61
12
0018-9456
Citations 
PageRank 
References 
1
0.36
1
Authors
6
Name
Order
Citations
PageRank
Wei Zhao194.55
Xiaodong Yang24613.17
Jiankang Xiao310.36
Qammer H. Abbasi411637.12
Hongbo Qin521.33
HuoRong Ren6203.69