Title
Fatigue Test Of Helical Nervous Electrodes And Weak Point Analysis Of Helical Nervous Electrodes Design
Abstract
Due to the increasing number of both implantation and removal of the helical nervous electrodes, the safety and the reliability of the electrode becomes an important issue in its clinical application E particularly its fatigue failure caused by body movement. Utilizing fatigue testing, we evaluated the weak points of the helical electrode. Our data analysis for fatigue cycles recorded by the fatigue test equipment showed that the adhesive strength between the silicone and suture, and the load of the electrode were essential for the mechanical durability of the electrode. The locations of the weak points and improvements for helical electrode design were given. The suggestions of decreasing damages to the nerves in clinical use will be helpful to the surgeons.
Year
DOI
Venue
2013
10.1109/EMBC.2013.6610959
2013 35TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC)
Keywords
Field
DocType
helical electrode, fatigue test, weak point
Biomedical engineering,Test equipment,Adhesion strength,Durability,Computer science,Fatigue testing,Silicone,Electrode
Conference
Volume
Issue
ISSN
2013
null
1557-170X
Citations 
PageRank 
References 
0
0.34
0
Authors
5
Name
Order
Citations
PageRank
Yuan Yuan100.34
Hongwei Hao223.57
Xiongwei Wen300.68
Xiaolong Mo400.34
Luming Li5117.60