Abstract | ||
---|---|---|
We present results from an experimental system employed to study the impact of glucose content on measured transmission data. The measurements have been performed at frequencies between 50 GHz and 75 GHz using V-Band waveguides, and the measured samples consisted of saline solutions with glucose concentrations down to 0.025 wt%. The performed transmission measurements show that it is possible to establish an approximately linear relationship between the transmission coefficient and the glucose content in the samples. This linear dependence is also verified via simulations. The results are of interest for developing a miniaturized mobile glucose sensing system. |
Year | DOI | Venue |
---|---|---|
2014 | 10.1109/MOBIHEALTH.2014.7015975 | Wireless Mobile Communication and Healthcare |
Keywords | Field | DocType |
biomedical measurement,chemical sensors,chemical variables measurement,sugar,waveguides,V-band waveguides,glucose content,miniaturized mobile glucose sensing system,saline solutions,transmission coefficient,transmission measurements,V-band,glucose sensing,non-invasive,transmission,waveguides | Analytical chemistry,Experimental system,Glucose sensing,Waveguide,Transmission coefficient,Materials science,Saline solutions,V band | Conference |
Citations | PageRank | References |
0 | 0.34 | 0 |
Authors | ||
8 |
Name | Order | Citations | PageRank |
---|---|---|---|
Helena Cano García | 1 | 0 | 0.34 |
Ioannis Papadopoulos-Kelidis | 2 | 0 | 0.34 |
Ioannis Sotiriou | 3 | 0 | 1.01 |
George Palikaras | 4 | 0 | 0.34 |
Efthymios Kallos | 5 | 0 | 0.34 |
Panagiotis Kosmas | 6 | 66 | 15.02 |
Clive Parini | 7 | 0 | 0.34 |
Papadopoulos-Kelidis, I. | 8 | 0 | 0.34 |