Abstract | ||
---|---|---|
The work progress on a new current transformer test set at PTB is described. The operation range for the ratio based bridge is intended for frequencies from 16.7 Hz up to 20 kHz. Equal setups of magnetically shielded current comparators with primary windings for currents from 100 mA up to 5 A convert the secondary currents of the standard current transformer and of the transformer under test (DUT) into proportional voltages. The errors of the DUT are calculated from the subsequent measurement of the complex voltage ratio. At 50 Hz, the measuring system offers a basic accuracy in the order of less than ± 2 ppm or μrad respectively. |
Year | DOI | Venue |
---|---|---|
2014 | 10.1109/AMPS.2013.6656217 | IEEE T. Instrumentation and Measurement |
Keywords | Field | DocType |
magnetic shielding,operation range,proportional voltages,current comparator (cc),two-channel sampling,current transformers,current transformer (ct),transformer testing,measurement uncertainty,current transformer test set,primary windings,transformer under test,complex voltage ratio,secondary currents,physikalisch-technische bundesanstalt,ratio based bridge,magnetically shielded current comparators,tut,bridge circuits,current transformer,two-channel sampling.,current comparators,ct test set,wideband current transformer bridge,current comparator | Energy efficient transformer,Current transformer,Delta-wye transformer,Distribution transformer,Transformer effect,Electronic engineering,Flyback transformer,Engineering,Electrical engineering,Linear variable differential transformer,Isolation transformer | Journal |
Volume | Issue | ISSN |
63 | 10 | 0018-9456 |
ISBN | Citations | PageRank |
978-1-4673-5571-1 | 6 | 1.56 |
References | Authors | |
1 | 5 |
Name | Order | Citations | PageRank |
---|---|---|---|
Enrico Mohns | 1 | 47 | 18.56 |
Johann Meisner | 2 | 10 | 3.88 |
Roeissle, G. | 3 | 6 | 1.56 |
Seckelmann, M. | 4 | 6 | 1.56 |
M. Korejwo | 5 | 6 | 1.56 |