Evaluation of an image-based tracking workflow using a passive marker and resonant micro-coil fiducials for automatic image plane alignment in interventional MRI. |
Abstract | ||
---|---|---|
In this paper, an original workflow is presented for MR image plane alignment based on tracking in real-time MR images. A test device consisting of two resonant micro-coils and a passive marker is proposed for detection using image-based algorithms. Micro-coils allow for automated initialization of the object detection in dedicated low flip angle projection images; then the passive marker is tracked in clinical real-time MR images, with alternation between two oblique orthogonal image planes along the test device axis; in case the passive marker is lost in real-time images, the workflow is reinitialized. The proposed workflow was designed to minimize dedicated acquisition time to a single dedicated acquisition in the ideal case (no reinitialization required). First experiments have shown promising results for test-device tracking precision, with a mean position error of 0.79 mm and a mean orientation error of 0.24°. |
Year | DOI | Venue |
---|---|---|
2012 | 10.1109/EMBC.2012.6346123 | EMBC |
Keywords | Field | DocType |
image-based tracking workflow,interventional mri,test-device tracking precision,image-based algorithms,real-time mr image,mean position error,resonant microcoil fiducials,clinical real-time mr image,biomedical equipment,biomedical mri,oblique orthogonal image plane,test device axis,object detection,mean orientation error,mr image plane alignment,low flip angle projection images,automatic image plane alignment,single dedicated acquisition,acquisition time,medical image processing,passive marker | Object detection,Computer vision,Fiducial marker,Computer science,Image plane,Image processing,Flip angle,Artificial intelligence,Initialization,Interventional magnetic resonance imaging,Workflow | Conference |
Volume | ISSN | ISBN |
2012 | 1557-170X | 978-1-4577-1787-1 |
Citations | PageRank | References |
1 | 0.40 | 2 |
Authors | ||
6 |
Name | Order | Citations | PageRank |
---|---|---|---|
M Neumann | 1 | 1 | 0.40 |
Breton, E. | 2 | 2 | 1.20 |
L Cuvillon | 3 | 1 | 0.40 |
L Pan | 4 | 1 | 0.40 |
C H Lorenz | 5 | 1 | 0.40 |
M de Mathelin | 6 | 1 | 0.40 |