Use of an inductive divider for AC voltage ratio calibration
Abstract
This paper presents the calibration of an electronic voltage divider and precise instrument voltage transformers in the range up to 1 kV at 50 Hz frequency using inductive voltage dividers. We also describe a way to evaluate the difference between a standard inductive divider and a calibrated instrument voltage transformer or electronic voltage divider expressed as the ratio error and phase displacement. The paper also presents the construction of a standard decadic voltage divider with a range of input voltage (10×100) V and (10×10) V.
Year
DOI
Venue
2013
10.1109/I2MTC.2013.6555398
Instrumentation and Measurement Technology Conference
Keywords
Field
DocType
calibration,potential transformers,voltage dividers,ac voltage ratio calibration,calibrated instrument voltage transformer,electronic voltage divider calibration,frequency 50 hz,inductive voltage dividers,phase displacement,precise instrument voltage transformers,standard decadic voltage divider construction,inductive voltage divider,ratio error,windings,materials,magnetic circuits