Abstract | ||
---|---|---|
Derives approximate analytical expressions for the local impulse response and covariance of images reconstructed from fully three-dimensional (3-D) positron emission tomography (PET) data using maximum a posteriori (MAP) estimation. These expressions explicitly account for the spatially variant detector response and sensitivity of a 3-D tomograph. The resulting spatially variant impulse response a... |
Year | DOI | Venue |
---|---|---|
2000 | 10.1109/42.870259 | IEEE Transactions on Medical Imaging |
Keywords | Field | DocType |
Positron emission tomography,Image reconstruction,Image resolution,Image analysis,Smoothing methods,Energy resolution,Spatial resolution,Lesions,Detectors,Fourier transforms | Iterative reconstruction,Impulse response,Computer vision,Monte Carlo method,Imaging phantom,Gaussian,Artificial intelligence,Maximum a posteriori estimation,Image resolution,Mathematics,Covariance | Journal |
Volume | Issue | ISSN |
19 | 5 | 0278-0062 |
Citations | PageRank | References |
63 | 5.00 | 11 |
Authors | ||
2 |
Name | Order | Citations | PageRank |
---|---|---|---|
Jinyi Qi | 1 | 284 | 35.82 |
Richard M Leahy | 2 | 1768 | 295.29 |