Title | ||
---|---|---|
Development of a hydrogen peroxide sensor based on screen-printed electrodes modified with inkjet-printed Prussian blue nanoparticles. |
Abstract | ||
---|---|---|
A sensor for the simple and sensitive measurement of hydrogen peroxide has been developed which is based on screen printed electrodes (SPEs) modified with Prussian blue nanoparticles (PBNPs) deposited using piezoelectric inkjet printing. PBNP-modified SPEs were characterized using physical and electrochemical techniques to optimize the PBNP layer thickness and electroanalytical conditions for optimum measurement of hydrogen peroxide. Sensor optimization resulted in a limit of detection of 2 x 10(-7) M, a linear range from 0 to 4.5 mM and a sensitivity of 762 mu A.mM(-1).cm(-2) which was achieved using 20 layers of printed PBNPs. Sensors also demonstrated excellent reproducibility (<5% rsd). |
Year | DOI | Venue |
---|---|---|
2014 | 10.3390/s140814222 | SENSORS |
Keywords | Field | DocType |
hydrogen peroxide,Prussian blue nanoparticles,screen-printed electrode,inkjet printing | Detection limit,Nanotechnology,Analytical chemistry,Linear range,Engineering,Hydrogen peroxide,Electrochemistry,Ferrocyanide,Electrode,Nanoparticle,Prussian blue | Journal |
Volume | Issue | ISSN |
14 | 8.0 | 1424-8220 |
Citations | PageRank | References |
1 | 0.87 | 0 |
Authors | ||
5 |
Name | Order | Citations | PageRank |
---|---|---|---|
Stefano Cinti | 1 | 1 | 0.87 |
Fabiana Arduini | 2 | 1 | 1.21 |
Danila Moscone | 3 | 1 | 1.21 |
Giuseppe Palleschi | 4 | 1 | 1.21 |
Anthony J Killard | 5 | 1 | 0.87 |