Title
A comparison of five different algorithms for EEG signal analysis in artifacts rejection for monitoring depth of anesthesia.
Abstract
•Five different algorithms are compared for depth of anesthesia monitoring.•Four kind of common artifacts in EEG are modeled as simulation data.•Permutation entropy is more reliable in artifacts rejection and EEG bandwidth.•Permutation entropy can track the drug effect more correctly.•The combined index via ANN are better than single index.
Year
DOI
Venue
2016
10.1016/j.bspc.2015.10.010
Biomedical Signal Processing and Control
Keywords
Field
DocType
Electroencephalography,Artifacts,Depth of anesthesia,Artifacts consideration
Christian ministry,Computer vision,Depth of anesthesia,Artificial intelligence,Medical physics,Translational medicine,Medicine,Eeg signal analysis
Journal
Volume
ISSN
Citations 
25
1746-8094
2
PageRank 
References 
Authors
0.40
7
5
Name
Order
Citations
PageRank
Quan Liu18621.65
Yi-Feng Chen240.76
Shou-Zen Fan38211.48
Maysam F. Abbod422428.14
Jiann Shing Shieh522428.44