Design Principles and Dynamic Front End Reconfiguration for Low Noise EEG Acquisition With Finger Based Dry Electrodes.
Abstract
Dry electrodes are a convenient alternative to wet electrodes for electroencephalography (EEG) acquisition systems. Dry electrodes are subject to a higher amount of noise at the electrode scalp interface and these effects can be worsened due to non-ideal design or improper placement on the head. In this work, we investigate a popular dry electrode design based on a number of resistive `finger' sha...
Year
DOI
Venue
2015
10.1109/TBCAS.2015.2471080
IEEE Transactions on Biomedical Circuits and Systems
Keywords
Field
DocType
Electrodes,Impedance,Noise,Electroencephalography,Scalp,Fingers,Impedance measurement