Title
An FPGA-based architecture of DSC-SRI units specially for motion blind ultrasound systems.
Abstract
Digital scan conversion (DSC) and speckle reduction imaging (SRI) are basic back-end units of an ultrasound imaging system. In a B-mode ultrasonography (USG) system, DSC is necessary to display a two dimensional image of a tissue structure. Speckle noise, present in an USG system, affects the quality of images observed. Moreover, motion blurring is observed while examining non-stationery organs like the heart. To address these issues, a high speed VLSI design of the USG back-end unit is presented in this paper. The back-end is designed in such a way that it can be integrated with the front-end of the USG system. An FPGA platform is used to prototype these units. Massive usage of the parallel-pipelining technique has helped in achieving high throughput at low power. The DSC architecture operates at 14.66 MHz and is capable of scan converting 132 raw frames per second. Similarly, the SRI unit operates at 221 MHz and is capable of de-speckling images of size 640 × 480 at 698 fps. Therefore, it is possible to investigate objects without motion blur using these units. System level integration issues of the DSC–SRI units are also covered in this paper.
Year
DOI
Venue
2015
10.1007/s11554-012-0289-y
Journal of Real-time Image Processing
Keywords
Field
DocType
Ultrasound sonography, Digital scan conversion, Speckle reduction imaging, DSP-FPGA system, System integration
Computer vision,Scan conversion,Computer science,Field-programmable gate array,Motion blur,Frame rate,Artificial intelligence,Throughput,Speckle noise,Very-large-scale integration,System integration
Journal
Volume
Issue
ISSN
10
3
1861-8219
Citations 
PageRank 
References 
1
0.36
18
Authors
4
Name
Order
Citations
PageRank
Rakesh Biswas162.66
Kishor Sarawadekar2146.58
Srinivas Varna310.36