Title | ||
---|---|---|
Interference measurements and failure analysis of patch-clamp laboratory instrumentation for electrophysiology tests. |
Abstract | ||
---|---|---|
This work analyzes the patch-clamp technique, which is the gold standard for ion channel measurements in electrophysiology, as well as the required instrumentation and main interferences that may affect the measurement accuracy. Different sources of degradation are evaluated, such as failure in headstage cooling and pipette/electrode impedance compensation systems, electromagnetic interference (EM... |
Year | DOI | Venue |
---|---|---|
2018 | 10.1109/MIM.2018.8423746 | IEEE Instrumentation & Measurement Magazine |
Keywords | Field | DocType |
Electromagnetic inteference,Ions,Current measurement,Degradation,Voltage measurement,Interference | Coupling,System of measurement,Electromagnetic interference,Electronic engineering,Interference (wave propagation),Engineering,Accuracy and precision,EMI,Instrumentation,Peristaltic pump | Journal |