Title | ||
---|---|---|
Brain-Shift Correction with Image-Based Registration and Landmark Accuracy Evaluation. |
Abstract | ||
---|---|---|
We describe an algorithm and its implementation details for automatic image-based registration of intra-operative ultrasound to MRI for brain-shift correction during neurosurgery. It is evaluated on a public database of 22 surgeries for retrospective evaluation, with a particular focus on choosing the appropriate transformation model and designing the most meaningful evaluation strategy. The method succeeds in a fully automatic fashion in all cases, with an average landmark registration error for the rigid model of 1.75 mm. |
Year | DOI | Venue |
---|---|---|
2018 | 10.1007/978-3-030-01045-4_17 | Lecture Notes in Computer Science |
DocType | Volume | ISSN |
Conference | 11042 | 0302-9743 |