Availability and variations of cardiac activity in the case of measuring the bioimpedance of wrist
Abstract
The utilization of technique for monitoring pulse wave from wrist by using the electrical bioimpedance has been implemented in this paper. A four electrode printed circuit board was used to evaluate the influence of positioning of the electrodes, pressure on the electrodes and individual differences of several volunteers on the availability of pulse wave from wrist. The influence of positioning of the electrode relative to radial artery was verified experimentally, allowing 5 mm shift in both transverse directions. The increasing pressure on the electrode was found to give rise to the time varying change of impedance, caused by pulse wave. This result was confirmed in the case of measurements on several volunteers, proposing the convergence in time varying change of impedance above the pressure of 400 g.