Title | ||
---|---|---|
Super-resolution ultrasound imaging with Gaussian fitting method and plane wave transmission. |
Abstract | ||
---|---|---|
Super-resolution ultrasound (SR-US) imaging can achieve a ten-fold resolution improvement compared with the traditional ultrasound technique, which is important for the medical diagnosis and treatment. However, challenges remain in SR-US imaging. In this paper, on one hand, a Gaussian fitting method, derived from optical localization microscopy, is used to improve the imaging spatial resolution of the SR-US. On the other hand, a plane wave technique is also used in US imaging for improving the imaging speed of the SR-US. To evaluate the performance of the proposed method, the numerical simulation was performed based on a phantom model. The experimental results indicate that by the use of a Gaussian fitting location method, combined with a plane wave transmission technique, we can accurately image the movement of microbubble in the phantom at a high frame rate, compared to the conventional B-model imaging. Hence, the technique makes it possible to achieve fast SR-US imaging. |
Year | DOI | Venue |
---|---|---|
2018 | 10.1117/12.2292956 | Proceedings of SPIE |
Keywords | DocType | Volume |
Ultrasound imaging,super-resolution,Gaussian fitting,plane wave transmission,numerical simulation | Conference | 10578 |
ISSN | Citations | PageRank |
0277-786X | 0 | 0.34 |
References | Authors | |
0 | 6 |
Name | Order | Citations | PageRank |
---|---|---|---|
Yuexia Shu | 1 | 1 | 2.04 |
Minglei Lv | 2 | 1 | 1.37 |
Ying Liu | 3 | 0 | 1.35 |
Zhuang-zhi Yan | 4 | 14 | 8.28 |
Jiehui Jiang | 5 | 48 | 5.39 |
Xin Liu | 6 | 287 | 74.92 |