Title | ||
---|---|---|
Circuit Design Challenges of ADC for the Application in Multiple Physiological Parameters Detection System |
Abstract | ||
---|---|---|
Design challenges of ADC for the application in multiple physiological parameters detection system is discussed in this paper. Based on requirements in the physiological signal detection applications, design difficulties of these high precision Sigma-Delta ADCs, including multi-channel high effective resolution, ultra-low power and low channel mismatch, are analyzed with design tradeoff considerations. Results of a Sigma-Delta modulator for physiological parameters detection applications is given. The noise floor of the modulator is about -120dB, the SNR is 101.4dB, and ENOB is 16.55 bit, with the power consumption of 18 mW. |
Year | DOI | Venue |
---|---|---|
2019 | 10.1109/ASICON47005.2019.8983501 | 2019 IEEE 13th International Conference on ASIC (ASICON) |
Keywords | Field | DocType |
multiple physiological parameters detection system,physiological signal detection applications,design difficulties,multichannel high effective resolution,low channel mismatch,physiological parameters detection applications,high precision Sigma-Delta ADC,noise figure 120.0 dB,noise figure 101.4 dB,word length 16.55 bit,power 18.0 mW | Flight dynamics (spacecraft),Noise floor,Detection theory,Computer science,Circuit design,Communication channel,Effective number of bits,Electronic engineering,Modulation,Power consumption | Conference |
ISSN | ISBN | Citations |
2162-7541 | 978-1-7281-0736-3 | 0 |
PageRank | References | Authors |
0.34 | 0 | 3 |
Name | Order | Citations | PageRank |
---|---|---|---|
Ye Yuan | 1 | 0 | 0.34 |
Song Ma | 2 | 0 | 0.34 |
Yu-Hua Cheng | 3 | 2 | 5.85 |