Distortion of the Intracranial Pressure Waveform by Extraventricular Drainage System. |
Abstract | ||
---|---|---|
Objective: To investigate whether intracranial pressure (ICP) waveform measurements obtained from extraventricular drainage (EVD) systems are suitable for the calculation of intracranial elastance (ICE) or cerebrovascular pressure autoregulation (PAR) indices. Methods: The transfer characteristic of an EVD system is investigated by its step and frequency responses with focus on the low frequency (... |
Year | DOI | Venue |
---|---|---|
2021 | 10.1109/TBME.2020.3036283 | IEEE Transactions on Biomedical Engineering |
Keywords | DocType | Volume |
Catheters,Pressure measurement,Ice,Monitoring,Distortion,Biomedical measurement,Resonant frequency | Journal | 68 |
Issue | ISSN | Citations |
5 | 0018-9294 | 0 |
PageRank | References | Authors |
0.34 | 0 | 3 |
Name | Order | Citations | PageRank |
---|---|---|---|
Daniel Teichmann | 1 | 0 | 0.34 |
James C. Lynch | 2 | 0 | 1.69 |
Thomas Heldt | 3 | 2 | 9.54 |