A Coin-Sized Detector for Room-Temperature X-Ray Photons and Electron Spectroscopy
Abstract
The system design and characterization of a compact single-SDD (Silicon Drift Detector) module for X-ray and electron spectroscopy in a windowless, room-temperature, vacuum environment is presented. The measured SDD leakage current density at +22° C is only 31 pA/cm <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sub> . The module, characterized at room temperature with <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">55</sub> Fe X-rays, demonstrated state-of-the-art performance: an energy resolution of 136 eV FWHM at 5.9 keV has been obtained with 1 μs filter peaking time, and its count-rate capability has been tested up to 4.2 Mcps with a 32 ns filter peaking time while maintaining an energy resolution of less than 200 eV FWHM at 5.9 keV.
Year
DOI
Venue
2020
10.1109/ICECS49266.2020.9294886
2020 27th IEEE International Conference on Electronics, Circuits and Systems (ICECS)
Keywords
DocType
ISBN
coin-sized Detector,X-ray photons,electron spectroscopy,compact single-SDD module,Silicon Drift Detector,windowless room-temperature,vacuum environment,measured SDD leakage current density,room temperature,55 Fe X-rays,energy resolution,FWHM,electron volt energy 136.0 eV,electron volt energy 5.9 keV,electron volt energy 200.0 eV,temperature 293.0 K to 298.0 K
Conference
978-1-7281-6045-0
Citations 
PageRank 
References 
0
0.34
0
Authors
5
Name
Order
Citations
PageRank
Matteo Gugiatti100.34
Marco Carminati22510.25
Carlo Fiorini312.39
Giacomo Borghi400.34
Francesco Ficorella500.34