Title
Noise Evaluation System for Biosignal Sensors Using Pseudo-Skin and Helmholtz Coil
Abstract
A test equipment that can evaluate the contact resistance and amount of noise of biosignal sensors is proposed in this study. Biosignals such as EEG are easily masked by the noise derived from contact resistance, because of their feebleness. To realize a low-noise measurement, the contact resistance and the amount of noise of biosignal sensors need to be investigated accurately prior to measuring the biosignal. A pseudo-skin made of conductive rubber was employed in this study to evaluate the performance of the sensors objectively and efficiently, similar to actual working conditions. The proposed device is composed of the pseudo-skin, signal-generating circuits to imitate EEG signals and pseudo hum noise on a minute scale, and a coil to emit the pseudo hum noise into the test space. The contact resistance of the pseudo-skin can be compared to that of human foreheads. The experimental results indicated that the standard deviation of the contact resistance of the pseudo-skin is 1.70, which is smaller than that of the human foreheads, which is 6.64. This result demonstrated that the pseudo-skin is suitable for contact resistance evaluation of the biosignal sensors. In addition, the correlation between contact resistance and the amount of noise was evaluated to assess the validity of the system. The amount of noise obtained was 597.37 μVrms, 1063.09 μVrms, and 1694.04 μVrms for the conductive gels with contact resistances of 32.65 kΩ, 85.17 kΩ, and 405 kΩ, respectively. An increase in the amount of noise with the increasing contact resistance was observed from the results. Further improvement of the device and an intensive study of the evaluation method of noise are required in future, in order to establish an efficient evaluation method for biosignal sensors.
Year
DOI
Venue
2019
10.1109/ISMICT.2019.8743712
2019 13th International Symposium on Medical Information and Communication Technology (ISMICT)
Keywords
Field
DocType
noise evaluation,biosignal measurement,contact resistance
Contact resistance,Evaluation system,Electrical conductor,Electronic engineering,Electromagnetic coil,Engineering,Helmholtz coil,Biosignal,Electronic circuit,Standard deviation
Conference
ISSN
ISBN
Citations 
2326-828X
978-1-7281-2343-1
0
PageRank 
References 
Authors
0.34
1
8
Name
Order
Citations
PageRank
Misaki Inaoka100.34
Shintaro Izumi28231.56
Shusuke Yoshimoto33012.56
Toshikazu Nezu400.68
Yuki Noda501.01
Teppei Araki601.69
Takafumi Uemura701.69
Tsuyoshi Sekitani85116.89